Arca
Skip to content
Tools
Tools

Advia Cantaur And ACS:180 Insulin Calibrator

Calibrator, Secondary. 510(k) K021535, from Bayer Diagnostics Corp. (Tarrytown, NY). Product code JIT, class 2.

As of Nov 18, 2026, K021535 is an FDA 510(k) held by Bayer Diagnostics Corp., substantially equivalent on Jun 27, 2002 after 48 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 27, 2002
  • Review time: 48 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ACS:180 and ADVIA Centaur Insulin calibrators are used for calibrating the ACS:180 and ADVIA Centaur Insulin Immunoassays. The ACS:180 and ADVIA Centaur Insulin are sandwich, chemiluminescence immunoassay for the quantitative determination of Insulin in human serum for use on the automated analyzer marketed by Bayer Corporation.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K021535, newest first; each opens FDA's recall record
Apr 5, 2017Z-2085-2017Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance…Terminated
Apr 5, 2017Z-2086-2017Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance…Terminated

Record

Type
Traditional
Received
May 10, 2002
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1150

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files