Arca
Skip to content
Tools
Tools

Modification To 30 Extension Set

Set, Administration, Intravascular. 510(k) K021395, from Churchill Medical Systems, Inc. (Dover, NH). Product code FPA, class 2.

As of Nov 26, 2026, K021395 is an FDA 510(k) held by Churchill Medical Systems, Inc., substantially equivalent on May 13, 2002 after 11 days, with 18 recalls tied to it in FDA's recall database.

  • Review time: 11 days
  • Recalls: 18

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Churchill Medical Systems Extension Sets is an accessory device used to deliver medical fluids.

Recalls

Recalls FDA ties to K021395, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 20, 2017Z-1484-2017Incorrect priming volume printed on the device package.Terminated
Nov 18, 2016Z-0787-2017Incorrect priming volume is printed on the device package.Terminated
Jan 15, 2015Z-1473-2015, Z-1478-2015, Z-1468-2015, Z-1472-2015, Z-1469-2015, Z-1470-2015, Z-1467-2015, Z-1477-2015, Z-1471-2015, Z-1476-2015, Z-1474-2015, Z-1475-2015Potential for leaking caused by insufficient bond between needleless device and female luer.Terminated
Jan 7, 2014Z-0813-2014Defective component may cause extension sets to leak.Terminated
Nov 16, 2010Z-1325-2011, Z-1326-2011, Z-1330-2011Routine testing conducted to challenge the success of sterilization was not fully completed for sterile products.Terminated

Other FPA clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every FPA device

Record

Type
Special
Received
May 2, 2002
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files