Arca
Skip to content

Medpor Plus Orbital Volume Replacement Impant

Implant, Eye Sphere. 510(k) K021357, from Porex Surgical, Inc. (Newnan, GA). Product code HPZ, class 2.

As of Nov 17, 2026, K021357 is an FDA 510(k) held by Porex Surgical, Inc., substantially equivalent on Aug 5, 2002 after 98 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Aug 5, 2002
  • Review time: 98 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

MEDPOR Surgical Implants in block, sheet and preformed shapes are intended for augmentation or restoration in the craniofacial region. *Porex Surgical, Inc.* *Page i* *15 Dart Road • Newman, GA 30265-1017 USA • TEL 1-800-521-7321 • 1-678-479-1610 • FAX 1-678-423-1435* *DEPARTMENT OF HEALTH & HUMAN SERVICES-USA* *DEPARTMENT OF HEALTH & HUMAN SERVICES* *Public Health Service* *Food and Drug…

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K021357, newest first; each opens FDA's recall record
StartedReasonStatus
Mar 20, 2012Z-1435-2012The MEDPOR PLUS SST EZ 22mm sphere implants are not within specification of 22mm +/- 8, therefore implant is not easily released from the syringe.Terminated

Other HPZ Clearances

The newest 510(k)s for the same product code; see every HPZ device
  • EzyporK212741 · Fci (France Chirurgie Instrumentation) Sas · Jun 24, 2022

Record

Type
Traditional
Received
Apr 29, 2002
Review panel
Ophthalmic
Regulation
21 CFR 886.3320

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files