Fassire-Duval Telescopic Im System
Rod, Fixation, Intramedullary And Accessories. 510(k) K020885, from Pega Medical, Inc. (Quebec, CA). Product code HSB, class 2.
As of Nov 17, 2026, K020885 is an FDA 510(k) held by Pega Medical, Inc., substantially equivalent on Feb 6, 2003 after 325 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Feb 6, 2003
- Review time: 325 days
- Recalls: 4
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis implant is indicated as a temporary implant to aid in the healing of long diaphysis fractures, osteotomies, malunions and nonunions and to prevent further fractures in femur, tibia and humerus in pediatric patients suffering from Osteogenesis Imperfecta without disrupting the bone growth plate.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Retrograde Pediatric NailK980854 · Encore Orthopedics, Inc. · Aug 11, 1998
- Synthes Elastic Intramedullary Nail (Ein) SystemK971783 · Synthes (Usa) · Jul 18, 1997
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Fassier-Duval Telescoping IM SystemK241983 · Orthopediatrics Canada Ulc Dba Pega Medical · Feb 21, 2025
- Fassier-Duval Telescopic IM SystemK211292 · Pega Medical, Inc. · May 26, 2021
- Gap Endo-Exo Medullary SystemK160545 · Pega Medical, Inc. · Nov 23, 2016
- Gap Endo-Exo Medullary SystemK111232 · Pega Medical, Inc. · Jan 26, 2012
- Fassier-Duval Telscopic Im SystemK041393 · Pega Medical, Inc. · Aug 17, 2004
Recalls
Recalls FDA ties to K020885, newest first; each opens FDA's recall record| Apr 10, 2015 | Z-1546-2015The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading… | Terminated |
| Mar 5, 2014 | Z-1472-2014The outer package label contained the incorrect length of the device. There is a label on the outer package and an additional label on the product contained in the package. | Terminated |
| Feb 7, 2014 | Z-1448-2014The Male Components in this lot are made of material with lower strength than manufacturing specifications. | Terminated |
| Jan 29, 2010 | Z-1494-2011Possible breakage. Structural weakness has been detected on the Male Driver 3.2mm and 4.0mm sizes. In certain cases, this weakness could lead to a premature failure of the instrument if it is submitted to large… | Terminated |
Other HSB Clearances
The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device- Freedom Nail Ankle Fusion SystemK261189 · McGinley Orthopaedic Innovations · Oct 19, 2026
- Pediatric Nailing Platform FemurK261345 · OrthoPediatrics Corp. · Sep 24, 2026
- Pangea Femur Reconstruction SystemK261481 · Stryker GmbH · Jun 29, 2026
- AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02)K253080 · Suzhou and Science & Technology Development Corp. · Jun 18, 2026
- Precice Max System and Precice Ankle Salvage SystemK261651 · NuVasive Specialized Orthopedics · Jun 18, 2026
- Fule Metallic Lockable Intramedullary Nail SystemsK253155 · Beijing Fule Science & Technology Development Co., Ltd. · Jun 17, 2026
- HERA Interlocking Intramedullary Nailing SystemK253046 · ZheJiang Decans Medical Devices Co., Ltd. · Jun 16, 2026
- Fitbone® Transport And Lengthening SystemK260146 · Orthofix Srl · May 19, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 18, 2002
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3020
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files