Arca
Skip to content
Tools
Tools

Bipap Synchrony Ventilatory Support System With Bi-Flex

Ventilator, Continuous, Non-Life-Supporting. 510(k) K020777, from Respironics, Inc. (Murrysville, PA). Product code MNS, class 2.

As of Nov 18, 2026, K020777 is an FDA 510(k) held by Respironics, Inc., substantially equivalent on Mar 20, 2002 after 9 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 9 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Synchrony is intended to provide non-invasive ventilation in adult patients (>30 kg) for the treatment of respiratory insufficiency (a condition in which the patient can continue without ventilation for some period, such as overnight) or obstructive sleep apnea. The Synchrony may be used in the hospital or home.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K020777, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 15, 2007Z-1061-2008, Z-1088-2008Foreign material: Glass fragments may present in the plastic bag material used to ship components.Terminated

Other MNS clearances

The newest 510(k)s for the same product code; their median review was 269 days; see every MNS device

Record

Type
Special
Received
Mar 11, 2002
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files