Philips Pagewriter Touch Cardiograph, Model M5000A Series

Electrocardiograph. 510(k) K020708, from Philips Medical Systems, Inc. (Andover, MA). Product code DPS, class 2.

As of Nov 17, 2026, K020708 is an FDA 510(k) held by Philips Medical Systems, Inc., substantially equivalent on Apr 25, 2002 after 51 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 25, 2002
  • Review time: 51 days
  • Recalls: 2
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule-out causes for symptoms.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K020708, newest first; each opens FDA's recall record
Sep 12, 2007Z-0124-2008Incorrect display: When using the 16 lead Patient Interface Module (PIM), lead tracings of V3R and V4R are reversed in the extended lead modes of Pediatric and Balanced.Terminated
Oct 13, 2004Z-0063-05Software can generate printed ECG's that associate incorrect patient data with the waveformTerminated

Other DPS Clearances

The newest 510(k)s for the same product code; their median review was 241 days; see every DPS device

Record

Type
Abbreviated, third-party review
Received
Mar 5, 2002
Review panel
Cardiovascular
Regulation
21 CFR 870.2340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.