Arca
Skip to content
Tools
Tools

Inratio

Test, Time, Prothrombin. 510(k) K020679, from Hemosense, Inc. (Milpitas, CA). Product code GJS, class 2.

As of Nov 18, 2026, K020679 is an FDA 510(k) held by Hemosense, Inc., substantially equivalent on May 6, 2002 after 63 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 63 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Feb 28, 2002 PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (As required by 21 CFR 807.87(j)) (As required by 21 CFR 807.87(j)) 510(K) Number: Device Name: HemoSense INRatio PT Monitoring System The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR).

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K020679, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 11, 2016Z-2355-2016, Z-2356-2016Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or…Terminated
Dec 5, 2014Z-0880-2015, Z-0881-2015In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has…Terminated
Oct 11, 2011Z-0188-2012Alere has initiated a recall of the Professional Disposables International (PDI) alcohol prep products due to the potential presence of low levels of the microorganism Bacillus cereus.Terminated
Sep 7, 2005Z-0132-06A software problem that may cause the INRatio meter to incorrectly display 'INR>7.5' test message under a particular use condition.Terminated

Other GJS clearances

The newest 510(k)s for the same product code; their median review was 299 days; see every GJS device

Record

Type
Traditional
Received
Mar 4, 2002
Review panel
Hematology
Regulation
21 CFR 864.7750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.