Super Secur-Fit Ha Hip Stems and Super Secur-Fit Plus Ha Hip Stems
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate. 510(k) K020615, from Howmedica Osteonics Corp. (Allendale, NJ). Product code MEH, class 2.
As of Nov 21, 2026, K020615 is an FDA 510(k) held by Howmedica Osteonics Corp., substantially equivalent on May 13, 2002 after 77 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 77 days
- Recalls: 8
- Cited as a predicate by: 1 device
- Secur-Fit Ha and Secur-Fit Plus Hip StemsK041170 · Howmedica Osteonics Corp. · Nov 9, 2004
| Jan 14, 2009 | Z-0557-2010, Z-0556-2010Stryker Orthopaedics became aware that there may be difficulty in mating stem to implantation/extraction instrument. | Terminated |
| Jan 6, 2009 | Z-1385-2009, Z-1387-2009, Z-1384-2009, Z-1680-2009, Z-1386-2009, Z-1383-2009Product deviation : Specific lots of sprayed HA hip stems did not meet Stryker's Internal Material Specification for tensile bond strength and crystallinity. | Terminated |
- Icona Hip StemK261576 · Corin, Ltd. · Aug 10, 2026
- Allure Hip Stem and Intramedullary PlugsK250375 · Waldemar Link GmbH & Co. KG · Nov 5, 2025
- Mfinity Femoral SystemK251166 · Medacta International S.A. · Jun 13, 2025
- Trivicta® Hip StemK251052 · Ortho Development Corp. · May 22, 2025
- Resolve Modular Revision Hip StemK242315 · United Orthopedic Corporation · May 1, 2025
- Alteon® HA Femoral StemsK243839 · Exactech, Inc. · Mar 10, 2025
- World LinerK243162 · Signature Orthopaedics Pty, Ltd. · Nov 25, 2024
- Icona Hip StemK241472 · Corin U.S.A. Limited · Oct 3, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Feb 25, 2002
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3353
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files