Unit, Phacofragmentation. 510(k) K020527, from American Optisurgical, Inc. (Lake Forest, CA). Product code HQC, class 2.

As of Nov 28, 2026, K020527 is an FDA 510(k) held by American Optisurgical, Inc., substantially equivalent on Jul 9, 2002 after 140 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 140 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K020527, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Mar 24, 2009Z-1808-2009Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that…Terminated

Other HQC clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every HQC device

Record

Type
Traditional
Received
Feb 19, 2002
Review panel
Ophthalmic
Regulation
21 CFR 886.4670

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files