Portable X-Ray Units and 1 more
System, X-Ray, Mobile. 510(k) K020436, from Sedecal., Sa. (Deer Field, IL). Product code IZL, class 2.
As of Nov 19, 2026, K020436 is an FDA 510(k) held by Sedecal., Sa., substantially equivalent on Feb 22, 2002 after 14 days, with 1 recall tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 1
- Cited as a predicate by: 1 device
Models SP-HF-2.8 and SP-HF-4.0 Portable X-ray Units are intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays.
- SedecalK103522 · Sedecal., Sa. · Jan 26, 2011
| Started | Recall | Status |
|---|---|---|
| Oct 27, 2003 | Z-0001-04A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard. | Terminated |
- EUROPA85, EUROPA85L, and EUROPA90 Portable X-ray Systems (EUROPA85, EUROPA85L, EUROPA90)K262028 · Aspen Imaging Healthcare, Inc. · Nov 6, 2026
- Remex-T(K)100CK261943 · Remedi, Inc. · Oct 2, 2026
- XliteK260236 · Gemss Healthcare Co., Ltd. · Aug 3, 2026
- Portable DR Imaging SystemK253058 · Shantou Institute of Ultrasonic Instruments Co., Ltd. · Jun 18, 2026
- RAYMO Mobile X-ray System (Model: RAYMO)K260085 · DRGEM Corporation · May 14, 2026
- Lux HD Detectors, DRX-LC Detector, DRX-Revolution Mobile X-Ray SystemK253185 · Carestream Health · Mar 11, 2026
- AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F (AirRay-mini)K253244 · Poskom Co., Ltd. · Feb 23, 2026
- Airbile-100K250976 · Poskom Co., Ltd. · Dec 19, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Feb 8, 2002
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1720
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files