Forex Corporation, Optima, Spinal System
Orthosis, Spinal Pedicle Fixation. 510(k) K020279, from Forex Corp. (Murray, UT). Product code MNI, class 2.
As of Nov 16, 2026, K020279 is an FDA 510(k) held by Forex Corp., substantially equivalent on Jun 20, 2002 after 143 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 20, 2002
- Review time: 143 days
- Recalls: 4
- Cited as a predicate by: 9 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe FOREX Corporation, OPTIMA™ posterior spinal fixation device is a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Amt Spinal SystemK002059 · Micron Precision Engineering, Inc. · Nov 13, 2000
- Xia Spine System HooksK001319 · Howmedica Osteonics Corp. · May 23, 2000
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- Diplomat® Spinal SystemK151704 · Signus Medizintechnik GmbH · Apr 13, 2016
- Palladian Lumbar Pedicle Screw SystemK143230 · Neurostructures, Inc. · Feb 25, 2015
- Pgs 5.5 Mm Cannulated Pedicle ScrewK141282 · Pinnacle Spine Group · Dec 2, 2014
- Medyssey Co., Ltd., Zenius Spinal SystemK093104 · Medyssey Co, Ltd. · Jun 25, 2010
- Zimmer Dto ImplantK071879 · Zimmer Spine, Inc. · Oct 5, 2007
- Delta Spinal Fusion SystemK071857 · Jemo Spine, LLC · Aug 8, 2007
- Lowtop Pedicle Screw SystemK070933 · Sanacor · Jun 13, 2007
- U&I, Optima, Spinal SystemK031585 · U&I Corp., America · Jun 27, 2003
- Optima, Spinal SystemK024096 · U&I Corp., America · Mar 12, 2003
Recalls
Recalls FDA ties to K020279, newest first; each opens FDA's recall record| Aug 10, 2004 | Z-1438-04Threads on a set screw, as a component of a spinal fixation system, may break off during use. | Terminated |
| Jan 14, 2004 | Z-0559-04Reports of hex driver tips breaking off or becoming worn or stripped. | Terminated |
| Jan 14, 2004 | Z-0558-04Reports of hex driver tips breaking off or becoming worn or stripped. | Terminated |
| Aug 18, 2003 | Z-1182-03Reports of pieces chipping off the tip of screwdriver during use. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 28, 2002
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files