Arca
Skip to content
Tools
Tools

Complete Brand Multi-Purpose Solution

Accessories, Soft Lens Products. 510(k) K014202, from Allergan, Inc. (Irvine, CA). Product code LPN, class 2.

As of Nov 19, 2026, K014202 is an FDA 510(k) held by Allergan, Inc., substantially equivalent on Feb 20, 2002 after 61 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 61 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

TRADE NAME Multi-Purpose Solution is indicated for the care of soft (hydrophilic) contact lenses. Use this product, as recommended by your eye care practitioner, to: • Chemically (NOT HEAT) Disinfect - Clean - Rinse - Store - Remove Protein

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K014202, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 23, 2010Z-0129-2011The recall was initiated after AMO determined, through internal testing, that a limited number of the flip top caps used during production of these solutions may leak and, although unlikely, the sterility of the product…Terminated
Sep 15, 2006Z-0345-2007Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm's plant…Terminated

Other LPN clearances

The newest 510(k)s for the same product code; their median review was 163 days; see every LPN device

Record

Type
Traditional
Received
Dec 21, 2001
Review panel
Ophthalmic
Regulation
21 CFR 886.5928

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files