Arca
Skip to content
Tools
Tools

Modification To: EUB-6500 Diagnostic Ultrasound Scanner

Transducer, Ultrasonic, Diagnostic. 510(k) K013723, from Hitachi Medical Corp. (Tarrytown, NY). Product code ITX, class 2.

As of Nov 20, 2026, K013723 is an FDA 510(k) held by Hitachi Medical Corp., substantially equivalent on Feb 6, 2002 after 89 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 89 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K013723, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 12, 2008Z-2054-2009The firm's parent company discovered a DICOM UID (Universal Identifier) code mix-up which can occur due to a software error. This error can cause the images from two (2) different patient scans to become intermingled…Terminated
May 7, 2004Z-1005-04Due to a software anomaly, various patient images may be intermixed within a single patient folder, which may result in mis-diagnosis and delay in treatment.Terminated

Other ITX clearances

The newest 510(k)s for the same product code; their median review was 178 days; see every ITX device

Record

Type
Special
Received
Nov 9, 2001
Review panel
Radiology
Regulation
21 CFR 892.1570

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files