Arca
Skip to content
Tools
Tools

Lightspeed 4.0 CT Scanner System

System, X-Ray, Tomography, Computed. 510(k) K013561, from GE Medical Systems (New Berlin, WI). Product code JAK, class 2.

As of Nov 21, 2026, K013561 is an FDA 510(k) held by GE Medical Systems, substantially equivalent on Nov 8, 2001 after 13 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 13 days
  • Recalls: 5
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K013561, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 10, 2016Z-1048-2016GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.Terminated
Jul 17, 2015Z-2801-2015Potential for expulsion of the Power Supply on Lightspeed CT scanner or Discovery ST PET CT or Discovery LS PET CT scanner.Terminated
Apr 22, 2015Z-1623-2015GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the…Terminated
Jul 25, 2011Z-1188-2013Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil…Terminated
Dec 8, 2003Z-0306-04Incorrect mounting hardware may cause X-ray tube to separate from the gantry and fall.Terminated

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional, third-party review
Received
Oct 26, 2001
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files