Prosthesis, Elbow, Constrained, Cemented. 510(k) K013042, from Biomet Orthopedics, Inc. (Warsaw, IN). Product code JDC, class 2.

As of Nov 20, 2026, K013042 is an FDA 510(k) held by Biomet Orthopedics, Inc., substantially equivalent on Oct 10, 2001 after 30 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 3
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Discovery Elbow is indicated for use with the following conditions: ■ Non-Inflammatory Degenerative Joint Disease including osteoarthritis and avascular necrosis ■ Rheumatoid arthritis ■ Revisions where other treatments or devices have failed ■ Correction of severe functional deformity - Treatment of acute or chronic fractures with humeral epicondyle involvement which are unmanageable using…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K013042, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 15, 2014Z-1925-2014Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.Terminated
May 23, 2008Z-0032-2009, Z-0033-2009The component in the package is not the correct size.Terminated

Record

Type
Special
Received
Sep 10, 2001
Review panel
Orthopedic
Regulation
21 CFR 888.3150

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.