V*Care Manual Resuscitator

Ventilator, Emergency, Manual (Resuscitator). 510(k) K012842, from Ventlab Corp. (Mocksville, NC). Product code BTM, class 2.

As of Nov 24, 2026, K012842 is an FDA 510(k) held by Ventlab Corp., substantially equivalent on Sep 18, 2001 after 26 days, with 21 recalls tied to it in FDA's recall database.

  • Review time: 26 days
  • Recalls: 21
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The V*Care Manual Resuscitator with Built-in Pressure Monitor's intended use is to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K012842, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 11, 2023Z-0367-2024, Z-0362-2024, Z-0368-2024, Z-0366-2024, Z-0361-2024, Z-0369-2024, Z-0364-2024, Z-0365-2024, Z-0363-2024A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.Open
May 16, 2014Z-1948-2014The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.Terminated
Jul 11, 2012Z-0177-2013, Z-0174-2013, Z-0176-2013, Z-0178-2013, Z-0180-2013, Z-0170-2013, Z-0173-2013, Z-0181-2013, Z-0171-2013, Z-0175-2013, Z-0179-2013Possible volume leakage through the inlet valve during compressions of the resuscitator bag.Terminated

Other BTM clearances

The newest 510(k)s for the same product code; their median review was 205 days; see every BTM device

Record

Type
Special
Received
Aug 23, 2001
Review panel
Anesthesiology
Regulation
21 CFR 868.5915

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files