Monoject Safety Needle

Needle, Hypodermic, Single Lumen. 510(k) K012736, from Tyco Healthcare (Mansfield, MA). Product code FMI, class 2.

As of Nov 18, 2026, K012736 is an FDA 510(k) held by Tyco Healthcare, substantially equivalent on Oct 25, 2001 after 70 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 25, 2001
  • Review time: 70 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The primary function of the device is to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The needle stick prevention feature of the device helps prevent accidental needle sticks by shielding the needle after use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K012736, newest first; the 25 newest; each opens FDA's recall record
Feb 9, 2021Z-1252-2021Safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and there is the potential for exposure to bloodborne pathogensTerminated
Oct 26, 2007Z-0729-2008Incorrect Needle size - Some of the kits contain a 1 inch 21G needle instead of the 1-1/2 inch 21G needle specified on the kit label.Terminated
Oct 26, 2007Z-0730-2008Incorrect Needle size - Some of the kits contain a 1 inch 21G needle instead of the 1-1/2 inch 21 needle specified on the kit label.Terminated
Oct 8, 2004Z-0146-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0102-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0068-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0143-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0099-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0074-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0153-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0126-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0140-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0096-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0079-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0105-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0065-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0086-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0119-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0087-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0098-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0120-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0112-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0130-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0071-05Needle cannula may separate from the hub during useTerminated
Oct 8, 2004Z-0104-05Needle cannula may separate from the hub during useTerminated

Other FMI Clearances

The newest 510(k)s for the same product code; their median review was 107 days; see every FMI device

Record

Type
Traditional
Received
Aug 16, 2001
Review panel
General Hospital
Regulation
21 CFR 880.5570

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files