Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker
Dc-Defibrillator, Low-Energy, (Including Paddles). 510(k) K012274, from Medtronic Physio-Control Corp. (Redmond, WA). Product code LDD, class 2.
As of Nov 20, 2026, K012274 is an FDA 510(k) held by Medtronic Physio-Control Corp., substantially equivalent on Feb 5, 2002 after 201 days, with 17 recalls tied to it in FDA's recall database.
- Review time: 201 days
- Recalls: 17
- Cited as a predicate by: 5 devices
Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
- Modified Lifepak 9 Monitor/DefibrillatorK902288 · Physio-Control Corp. · Feb 25, 1991
- Life Pak 12K973486 · Physio-Control Corp. · Jan 9, 1998
- Lifepak 12 BiphasicK991910 · Physio-Control Corp. · Sep 3, 1999
- Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorK130454 · Physio-Control, Inc. · Aug 21, 2013
- Lifepak 20EK073089 · Physio-Control, Inc. · Jul 29, 2008
- Lifepak 20 Defibrillator/ MonitorK063119 · Medtronic Emergency Response Systems, Inc. · Dec 22, 2006
- Powerheart EcdK052316 · Cardiac Science, Inc. · Jan 3, 2006
- Lifepak 12, 20, 500, CR PlusK033275 · Medtronic Physio-Control Corp. · Nov 6, 2003
| May 27, 2010 | Z-1903-2010A failure on the power supply assembly can result in either "No DC power" or "No DC or AC power." A failure of DC (battery) power can result in a delay of defibrillation therapy if no AC (line) power is available and… | Terminated |
| May 27, 2010 | Z-1722-2011A component failure on the AC power supply assembly results in "NO AC POWER" operation. A failure of AC power can result in a delay of defibrillation therapy if backup battery power is depleted. | Terminated |
| Dec 30, 2008 | Z-0185-2012When users do not disconnect the test plug and reconnect the QUIK-COMBO electrodes to the therapy cable, the test plug signal can be misinterpreted as a patient waveform. | Terminated |
| Jun 18, 2008 | Z-2388-2008Potential for the coin battery to drain prematurely causing the monitor clock time and date to be incorrect and the service light indicator to illuminate. | Terminated |
| Jan 8, 2008 | Z-1050-2008Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA). | Terminated |
| Apr 27, 2007 | Z-0925-2007White Screen Lock up: LifePak 20 defibrillators with software version 48 or version 52 may experience an intermittent timing issue during the power on sequence. This may lock up the device resulting in the monitor… | Terminated |
| Mar 30, 2007 | Z-0886-2007Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power. | Terminated |
| Aug 30, 2006 | Z-1502-06Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning. | Terminated |
| Mar 22, 2006 | Z-0711-06Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions. | Terminated |
| Mar 16, 2006 | Z-0295-2007LIFEPAK 20 may lock-up when attempting to power-up on DC within 2 seconds after removing AC power. | Terminated |
| Sep 2, 2005 | Z-0117-06, Z-0116-06Potential for voltage pins in connector of therapy cable to break. | Terminated |
| Jun 17, 2005 | Z-0984-05The defibrillator may not operate if the device is turned on during an AC Loss Alert. | Terminated |
| May 5, 2004 | Z-1107-04Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in… | Terminated |
| May 5, 2004 | Z-0971-04The DC Power Adapter may fail to charge LIFEPAK NiCd 2.4 AH batteries used with the LIFEPAK 12 defibrillator/monitor. | Terminated |
| Apr 5, 2004 | Z-0938-04Potential for display to go blank due to an electrical component that may fail due to mechanical shock of vibration of the defibrillator. | Terminated |
| Nov 12, 2003 | Z-0294-04Device unable to recognize the connection to the hard paddles accessory. | Terminated |
- Switched Internal PaddlesK203231 · Philips North America, LLC · Jan 11, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 19, 2001
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5300
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files