Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker

Dc-Defibrillator, Low-Energy, (Including Paddles). 510(k) K012274, from Medtronic Physio-Control Corp. (Redmond, WA). Product code LDD, class 2.

As of Nov 20, 2026, K012274 is an FDA 510(k) held by Medtronic Physio-Control Corp., substantially equivalent on Feb 5, 2002 after 201 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 201 days
  • Recalls: 17
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K012274, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 27, 2010Z-1903-2010A failure on the power supply assembly can result in either "No DC power" or "No DC or AC power." A failure of DC (battery) power can result in a delay of defibrillation therapy if no AC (line) power is available and…Terminated
May 27, 2010Z-1722-2011A component failure on the AC power supply assembly results in "NO AC POWER" operation. A failure of AC power can result in a delay of defibrillation therapy if backup battery power is depleted.Terminated
Dec 30, 2008Z-0185-2012When users do not disconnect the test plug and reconnect the QUIK-COMBO electrodes to the therapy cable, the test plug signal can be misinterpreted as a patient waveform.Terminated
Jun 18, 2008Z-2388-2008Potential for the coin battery to drain prematurely causing the monitor clock time and date to be incorrect and the service light indicator to illuminate.Terminated
Jan 8, 2008Z-1050-2008Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).Terminated
Apr 27, 2007Z-0925-2007White Screen Lock up: LifePak 20 defibrillators with software version 48 or version 52 may experience an intermittent timing issue during the power on sequence. This may lock up the device resulting in the monitor…Terminated
Mar 30, 2007Z-0886-2007Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.Terminated
Aug 30, 2006Z-1502-06Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.Terminated
Mar 22, 2006Z-0711-06Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.Terminated
Mar 16, 2006Z-0295-2007LIFEPAK 20 may lock-up when attempting to power-up on DC within 2 seconds after removing AC power.Terminated
Sep 2, 2005Z-0117-06, Z-0116-06Potential for voltage pins in connector of therapy cable to break.Terminated
Jun 17, 2005Z-0984-05The defibrillator may not operate if the device is turned on during an AC Loss Alert.Terminated
May 5, 2004Z-1107-04Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in…Terminated
May 5, 2004Z-0971-04The DC Power Adapter may fail to charge LIFEPAK NiCd 2.4 AH batteries used with the LIFEPAK 12 defibrillator/monitor.Terminated
Apr 5, 2004Z-0938-04Potential for display to go blank due to an electrical component that may fail due to mechanical shock of vibration of the defibrillator.Terminated
Nov 12, 2003Z-0294-04Device unable to recognize the connection to the hard paddles accessory.Terminated

Other LDD clearances

The newest 510(k)s for the same product code; see every LDD device

Record

Type
Traditional
Received
Jul 19, 2001
Review panel
Cardiovascular
Regulation
21 CFR 870.5300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files