Arca
Skip to content

Synthes (USA) Trochanteric Fixation Nail (Tfn) System

Rod, Fixation, Intramedullary And Accessories. 510(k) K011857, from Synthes (Usa) (Paoli, PA). Product code HSB, class 2.

As of Nov 17, 2026, K011857 is an FDA 510(k) held by Synthes (Usa), substantially equivalent on Sep 12, 2001 after 90 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 12, 2001
  • Review time: 90 days
  • Recalls: 7
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

k011857 Synthes (USA) Trochanteric Fixation Nail System Synthes Trochanteric Fixation Nail (TFN) System is intended to treat stable and unstable fractures of the proximal femur including pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations thereof.

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K011857, newest first; each opens FDA's recall record
Jul 21, 2017Z-0604-2018The locking mechanism and protective cap were missing from the sterile packed nails of the affected lot.Terminated
Nov 2, 2015Z-0375-2016TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling…Terminated
Oct 29, 2014Z-0219-2015The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise…Terminated
Jul 31, 2013Z-0387-2014The Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and became stuck.Terminated
Jul 30, 2013Z-0474-2014Certain lots of Trochanteric Fixation Nail were distributed to the field with the incorrect labeling of the length and diameter of the nail on the packaging.Terminated
Jul 30, 2013Z-0452-2014Two lots of Titanium Trochanteric Fixation Nail (TFN) System are missing the second bend in the nail.Terminated
Nov 2, 2012Z-2058-2013Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be…Terminated

Other HSB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device

Record

Type
Traditional
Received
Jun 14, 2001
Review panel
Orthopedic
Regulation
21 CFR 888.3020

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.