Neurological Stereotaxic Instrument. 510(k) K011775, from FHC, Inc. (Bowdoinham, ME). Product code HAW, class 2.
As of Nov 19, 2026, K011775 is an FDA 510(k) held by FHC, Inc., substantially equivalent on Aug 30, 2001 after 84 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 8
- Cited as a predicate by: 1 device
- Microtargeting Drive SystemK003776 · FHC, Inc. · Feb 23, 2001
- Cosman Robert Wells Functional Probe MicrodriveK992721 · Radionics, Inc. · Sep 10, 1999
- Axon MP-1 MicropositionerK990683 · Axon Instruments, Inc. · Nov 3, 1999
- Microelectroencephalography Professional System 5000 (Eeg Pro System 5000)K991077 · Microrecording Systems Consultants, LLC · Jun 9, 1999
- Elekta Microdrive, Model 916716K040979 · Elekta Instrument AB · Jul 23, 2004
| Oct 4, 2024 | Z-0312-2025A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the… | Open |
| May 10, 2018 | Z-2348-2018, Z-2355-2018, Z-2345-2018, Z-2346-2018, Z-2347-2018, Z-2356-2018The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple… | Terminated |
| Mar 14, 2005 | Z-0895-05Potential for non-sterile pin to contaminate sterile field | Terminated |
- Stereotactic Ultralight System SUSyK261157 · inomed Medizintechnik GmbH · Nov 6, 2026
- StealthStation™ S8 Cranial v2.3.0K262716 · Medtronic Navigation · Nov 6, 2026
- NeuroAlignK261997 · Medivis, Inc. · Oct 26, 2026
- Disposable Reflective Marker Spheres (AM-BK115S, AM-BL130S)K262214 · Guangzhou Aimooe Technology Co., Ltd. · Oct 15, 2026
- Geniant Cranial (Navigated Neurosurgical Positioning Robot)K253856 · Koh Young Technology, Inc. · Sep 10, 2026
- Leksell® Vantage Stereotactic SystemK261932 · Elekta Solutions AB · Aug 5, 2026
- Zeta Navigation SystemK253663 · Zeta Surgical, Inc. · May 1, 2026
- Stealth AXiS Cranial clinical applicationK253379 · Medtronic Navigation, Inc. · Mar 26, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 7, 2001
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files