Powered Laser Surgical Instrument. 510(k) K011703, from Lumenis (Santa Clara, CA). Product code GEX, class 2.

As of Nov 18, 2026, K011703 is an FDA 510(k) held by Lumenis, substantially equivalent on Aug 29, 2001 after 89 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 29, 2001
  • Review time: 89 days
  • Recalls: 2
  • Cited as a predicate by: 15 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The modified and the currently marketed VersaPulse PowerSuite Holmium (Ho:YAG) and Dual Wavelength (Ho:YAG/Nd:YAG) Surgical Lasers and Delivery Devices with accessories are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: - Ho:YAG - urology; urinary…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

15 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K011703, newest first; each opens FDA's recall record
Dec 18, 2007Z-0585-2008Incorrect Label: The Units outer pouch is labeled 200 micro fiber however, the inner pouch's label affixed to the product indicated that it is a 365 micron fiber. The micron fiber contained in the inner pouch is 200…Terminated
Aug 29, 2006Z-0028-2007Device lacks the black indicator markings on the metal tip and there is the potential for fiber degradation (detachment of the metal cap, and the fiber lasing straight) during use.Terminated

Other GEX Clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every GEX device

Record

Type
Traditional
Received
Jun 1, 2001
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4810

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.