Arca
Skip to content
Tools
Tools

Neurogen Nerve Guide

Cuff, Nerve. 510(k) K011168, from Integra Lifesciences Corp. (Plainsboro, NJ). Product code JXI, class 2.

As of Nov 20, 2026, K011168 is an FDA 510(k) held by Integra Lifesciences Corp., substantially equivalent on Jun 22, 2001 after 66 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 66 days
  • Recalls: 2
  • Cited as a predicate by: 8 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

NeuroGen™ Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.

Cited as a predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K011168, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 7, 2020Z-1470-2020Nonconforming product which was dispositioned for rejection was mistakenly released for commercial use.Terminated
Mar 28, 2013Z-1287-2013Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.Terminated

Other JXI clearances

The newest 510(k)s for the same product code; their median review was 105 days; see every JXI device

Record

Type
Traditional
Received
Apr 17, 2001
Review panel
Neurology
Regulation
21 CFR 882.5275

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.