Arrow-Trerotola Percutaneous Thrombolytic Device Or Ptd
Peripheral Mechanical Thrombectomy With Aspiration. 510(k) K011056, from Arrow Intl., Inc. (Reading, PA). Product code QEW, class 2.
As of Nov 20, 2026, K011056 is an FDA 510(k) held by Arrow Intl., Inc., substantially equivalent on Oct 24, 2001 after 201 days, with 13 recalls tied to it in FDA's recall database.
- Review time: 201 days
- Recalls: 13
- Cited as a predicate by: 2 devices
- Arrow-Trerotola Percutaneous Thrombolytic Device Or Ptd, Models PT-45509 (Catheter Kit With Introducer Sheaths),PT-65K990829 · Arrow Intl., Inc. · May 4, 1999
- Cleaner Rotational Thrombectomy System (9 MM X 65 Cm) and (9 MM X 120 Cm)K091029 · Rex Medical · Feb 18, 2010
- Akonya Eliminator Mechanical Thrombectomy Device, Model EL10060060K041374 · Idev Technologies, Inc. · Mar 2, 2005
| Jan 11, 2022 | Z-0628-2022, Z-0623-2022, Z-0626-2022, Z-0624-2022, Z-0627-2022, Z-0629-2022, Z-0625-2022The firm received reports indicating PTD tip separation during use. | Terminated |
| Sep 20, 2021 | Z-0231-2022, Z-0233-2022, Z-0232-2022, Z-0234-20227FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy. | Terminated |
| Feb 11, 2009 | Z-1502-2009Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism. | Terminated |
| Mar 18, 2005 | Z-0842-05The orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use. | Terminated |
- INDIGO® Aspiration SystemK262904 · Penumbra, Inc. · Oct 15, 2026
- Triever Duo CatheterK253419 · Inari Medical, Inc. · Sep 18, 2026
- Helo Max Thrombectomy System (HEL2090)K261800 · Endovascular Engineering · Sep 3, 2026
- Artix™ MT;Artix™ Thin-Walled SheathK262538 · Inari Medical, Inc. · Aug 21, 2026
- Vertex™ Thrombectomy SystemK261523 · Jupiter Endovascular · Aug 4, 2026
- INDIGO® Aspiration System – INDIGO LinkK260599 · Penumbra, Inc. · Apr 24, 2026
- INDIGO® Aspiration System - Lightning Flash Aspiration Tubing with LINKK251949 · Penumbra, Inc. · Feb 5, 2026
- CLEANER™ Vac Thrombectomy System, CLEANER™ Vac Aspiration Catheter with Handpiece, CLEANER™ Vac Aspiration CanisterK260028 · Argon Medical Devices, Inc. · Feb 5, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 6, 2001
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5150
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files