Arca
Skip to content
Tools
Tools

Oncology Salvage System

Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented. 510(k) K002757, from Biomet, Inc. (Warsaw, IN). Product code JDI, class 2.

As of Nov 21, 2026, K002757 is an FDA 510(k) held by Biomet, Inc., substantially equivalent on Nov 24, 2000 after 80 days, with 14 recalls tied to it in FDA's recall database.

  • Review time: 80 days
  • Recalls: 14
  • Cited as a predicate by: 21 devices

Cited as a predicate

21 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K002757, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 25, 2020Z-0623-2021Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.Terminated
Apr 17, 2020Z-2147-2020Potential presence of elevated endotoxin levels that exceed the specification limitTerminated
Oct 10, 2018Z-0590-2019Zimmer Biomet is conducting a lot specific medical device recall for two lots of the Orthopedic Salvage System (OSS) and the Comprehensive Shoulder System. The investigation determined that the Mini Humeral Stem is…Terminated
Jun 5, 2017Z-0303-2018Possibility that the patient label inside the product box was incorrectly labeled. The outer box label and actual product are correctly identified but the internal patient label could be incorrect.Terminated
Nov 1, 2011Z-0417-2012Biomet has initiated this action following an investigation which identified that Part 150367, (OSS Cemented 1M Stem 13x150) was incorrectly labeled as part 150368 (OSS Cemented 1M Stem 14x150).Terminated
May 19, 2009Z-1528-2009, Z-1530-2009, Z-1527-2009, Z-1529-2009, Z-1531-2009, Z-1526-2009The retaining ring may fall out of the instrument during surgery.Terminated
May 15, 2008Z-1881-2008, Z-1880-2008, Z-1879-2008Mislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while…Terminated

Other JDI clearances

The newest 510(k)s for the same product code; their median review was 219 days; see every JDI device

Record

Type
Traditional
Received
Sep 5, 2000
Review panel
Orthopedic
Regulation
21 CFR 888.3350

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.