Ecat Pet Scanner
System, Tomography, Computed, Emission. 510(k) K002715, from Cti Pet Systems, Inc. (Knoxville, TN). Product code KPS, class 2.
As of Nov 18, 2026, K002715 is an FDA 510(k) held by Cti Pet Systems, Inc., substantially equivalent on Oct 31, 2000 after 61 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 61 days
- Recalls: 2
- Cited as a predicate by: 5 devices
Cited as a predicate
5 later 510(k) summaries name this device as a predicate- Pem 2400 Pet ScannerK032063 · Pem Technologies · Aug 18, 2003
- Ecat Pet/Ct; Ecat Lso Pet/Ct; Ecat Lso Pet/Ct 16K023768 · Cti Pet Systems, Inc. · Nov 25, 2002
- Ecat Vision Pet Scanner, Models 3000 & 5000K021633 · Cti Pet Systems, Inc. · Aug 14, 2002
- Gemini Imaging SystemK013521 · Adac Laboratories · Nov 7, 2001
- Modification To: Ecat Lso Pet/Ct ScannerK013504 · Cti Pet Systems, Inc. · Oct 31, 2001
Recalls
Recalls FDA ties to K002715, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| May 27, 2004 | Z-0547-05Misalignment of CT and PET scans from performing additional CT scans may produce an incorrect fused data set. In performing additional CT scans, the original PET and CT scans may become misaligned and produce an… | Terminated |
| Oct 2, 2003 | Z-0106-04CPS has found that in some instances, when the CT E-Stop is activated, a malfunction may occur where the PHS controller may not shut down and stop bed motion. | Terminated |
Other KPS clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 31, 2000
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files