Exacta-Mix 2400 Compounding System, Model Exacta-Mix 2400

Set, I.V. Fluid Transfer. 510(k) K002705, from Baxa Corp. (Englewood, CO). Product code LHI, class 2.

As of Nov 20, 2026, K002705 is an FDA 510(k) held by Baxa Corp., substantially equivalent on Mar 28, 2001 after 210 days, with 15 recalls tied to it in FDA's recall database.

  • Review time: 210 days
  • Recalls: 15
  • Cited as a predicate by: 4 devices

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K002705, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 22, 2026Z-2262-2026Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system…Open
Aug 20, 2024Z-3085-2024, Z-3082-2024, Z-3084-2024, Z-3083-2024Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the…Open
Dec 22, 2023Z-1002-2024, Z-1001-2024An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the…Open
Jul 14, 2022Z-1570-2022There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.Open
Apr 25, 2018Z-1908-2018, Z-1909-2018, Z-1910-2018, Z-1911-2018Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used…Terminated
Jun 4, 2013Z-1752-2013Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight…Terminated
Feb 10, 2010Z-1197-2010Molding defect in the valve body causes leaking.Terminated
Nov 7, 2003Z-0210-04If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber..Terminated

Other LHI clearances

The newest 510(k)s for the same product code; their median review was 259 days; see every LHI device

Record

Type
Traditional
Received
Aug 30, 2000
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files