Ecat Accel Pet Scanner
System, Tomography, Computed, Emission. 510(k) K002584, from Cti Pet Systems, Inc. (Knoxville, TN). Product code KPS, class 2.
As of Nov 18, 2026, K002584 is an FDA 510(k) held by Cti Pet Systems, Inc., substantially equivalent on Nov 16, 2000 after 87 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 87 days
- Recalls: 14
- Cited as a predicate by: 3 devices
Cited as a predicate
3 later 510(k) summaries name this device as a predicateRecalls
Recalls FDA ties to K002584, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 23, 2007 | Z-0273-2008Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all… | Terminated |
| Jul 23, 2007 | Z-0248-2008, Z-0250-2008, Z-0249-2008, Z-0252-2008, Z-0247-2008, Z-0246-2008, Z-0251-2008Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix… | Terminated |
| Jul 23, 2007 | Z-0272-2008, Z-0270-2008, Z-0269-2008, Z-0271-2008Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all… | Terminated |
| Apr 5, 2005 | Z-0819-05Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit. | Terminated |
| Mar 4, 2003 | Z-0674-03Incorrect patient information is displayed in the reconstructed images of a whole body patient scan. | Terminated |
Other KPS clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device- Micsi-PetK261305 · Microstructure Imaging, Inc. · Sep 11, 2026
- uMI Panvivo (uMI Panvivo M)K261427 · Shanghai United Imaging Healthcare Co., Ltd. · Jul 28, 2026
- Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7 (11364753);Symbia Pro.specta VA40…K261315 · Siemens Medical Solutions USA, Inc. · May 21, 2026
- uMI Panvivo (uMI Panvivo LS); uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI…K260524 · Shanghai United Imaging Healthcare Co., Ltd. · Apr 3, 2026
- uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)K253564 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 13, 2026
- VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System…K254001 · Spectrum Dynamics Medical, Ltd. · Jan 13, 2026
- AnyScan 3.0 NM Scanner FamilyK253844 · Mediso Medical Imaging Systems, Ltd. · Dec 30, 2025
- Cartesion Prime (PCD-1000A/3) V10.21K251370 · Canon Medical Systems Corporation · Dec 1, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 21, 2000
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files