Thrombelastograph Coagulation Analyzer TEG-5000 Series
System, Multipurpose For In Vitro Coagulation Studies. 510(k) K002177, from Haemoscope Corp. (Skokie, IL). Product code JPA, class 2.
As of Nov 18, 2026, K002177 is an FDA 510(k) held by Haemoscope Corp., substantially equivalent on Sep 14, 2000 after 57 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 57 days
- Recalls: 7
- Cited as a predicate by: 4 devices
Cited as a predicate
4 later 510(k) summaries name this device as a predicate- TEG 6s Hemostasis System, TEG 6s Citrated: K, RT, FF Assay CartridgeK183160 · Haemonetics Corporation · May 9, 2019
- CORA (Coagulation Resonance Analysis) SystemK150041 · Coramed Technologies, LLC · Jun 26, 2015
- Ex-Tem. Fib-Tem, and Ap-Tem for Rotem Delta Thromboelastometry SystemK101533 · Tem Innovations GmbH · Aug 10, 2011
- Rotem Delta Thromboelastometry SystemK083842 · Pentapharm GmbH · Mar 23, 2010
Recalls
Recalls FDA ties to K002177, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 26, 2022 | Z-1310-2022Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference… | Terminated |
| Apr 19, 2022 | Z-1298-2022When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion… | Terminated |
| Feb 19, 2020 | Z-1617-2020Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded… | Terminated |
| Jan 26, 2015 | Z-1130-2015Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter. | Terminated |
| Jun 20, 2013 | Z-1796-2013Research Use Only (RUO) parameters were included in TEG 5000 User Manual. | Terminated |
| Mar 4, 2011 | Z-2742-2011Haemoscope received a report that during the process of Validation, Level II QC was running with an out-of-range low MA (Maximum Amplitude). | Terminated |
| Jan 28, 2011 | Z-1738-2011Haemonetics has received a report of smoke from the unit with a power supply failure. | Terminated |
Other JPA clearances
The newest 510(k)s for the same product code; their median review was 158 days; see every JPA device- Sonoclot Coagulation & Platelet Function Analyzer System with Sonoclot Viewer (SCP1); Sonoclot Coagulation & Platelet…K260909 · Sienco, Inc. · Oct 27, 2026
- STA Satellite Max®K253658 · Diagnostica Stago, Inc. · Apr 27, 2026
- Automated Blood Coagulation Analyzer CN-Series (CN-6000)K250965 · Sysmex America, Inc. · Jun 2, 2025
- TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeK251024 · Haemonetics · Apr 30, 2025
- TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeK243858 · Haemonetics Corporation · Jan 15, 2025
- Citrated: K, KH, RTH, FFHK232018 · Haemonetics Corporation · Mar 29, 2024
- Acl Top 970 CLK233790 · Instrumentation Laboratory · Dec 29, 2023
- ACL TOP 970 CL, HemosIL CL Anti-Cardiolipin IgM, HemosIL CL Anti-ß2 Glycoprotein-I IgMK221359 · Instrumentation Laboratory CO · Sep 29, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 19, 2000
- Review panel
- Hematology
- Regulation
- 21 CFR 864.5425
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files