Philips Integris Allura

Interventional Fluoroscopic X-Ray System. 510(k) K002016, from Philips Medical Systems North America, Inc. (Da Best, NL). Product code OWB, class 2.

As of Nov 19, 2026, K002016 is an FDA 510(k) held by Philips Medical Systems North America, Inc., substantially equivalent on Sep 6, 2000 after 65 days, with 13 recalls tied to it in FDA's recall database.

  • Review time: 65 days
  • Recalls: 13
  • Cited as a predicate by: 3 devices

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K002016, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 22, 2024Z-3167-2024Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.Open
Oct 4, 2023Z-0227-2024Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer)…Open
Dec 30, 2019Z-1487-2020, Z-1488-2020A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.Terminated
Nov 3, 2015Z-0727-2016, Z-0726-2016, Z-0729-2016, Z-0728-2016The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.Terminated
Aug 8, 2014Z-0022-2015Philips had discovered through a field service testing where a device failed to comply with a Federal standard.Terminated
Dec 14, 2006Z-0366-2007Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.Terminated
Oct 12, 2005Z-0206-06, Z-0207-06Potential for high voltage exposure involving the injector interface connector of patient support tables.Terminated
Jul 27, 2005Z-0022-061) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release…Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Jul 3, 2000
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files