System, Nuclear Magnetic Resonance Imaging. 510(k) K001334, from Hitachi Medical Systems America, Inc. (Twinsburg, OH). Product code LNH, class 2.
As of Nov 19, 2026, K001334 is an FDA 510(k) held by Hitachi Medical Systems America, Inc., substantially equivalent on Jul 6, 2000 after 70 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 70 days
- Recalls: 3
- Cited as a predicate by: 6 devices
- Oper Series Open Type Permanent-Magnet Mri System, Model: OPER-0.3, -0.35, -0.4K092899 · Ningbo Xingaoyi Magnetism Co., Ltd. · Nov 10, 2009
- I Open 0.5TK090873 · Beijing Wandong Medical Equipment Co., Ltd. · Apr 10, 2009
- I_Open 0.4TK090433 · Beijing Wandong Medical Equipment Co., Ltd. · Mar 25, 2009
- Airis Elite Magnetic Resonance Imaging DeviceK032232 · Hitachi Medical Systems America, Inc. · Sep 29, 2003
- Superopen 0.35T, Model NSM-P035K030918 · Neusoft Digital Medical Systems Co., Ltd. · Apr 1, 2003
- Hitachi AltaireK022390 · Hitachi Medical Systems America, Inc. · Oct 11, 2002
| Apr 19, 2010 | Z-1616-2010Hitachi determined that the mounting method for a Compensation Coil did not include an appropriate factor of safety, so it is possible that the coil could fall, contacting the patient table and potentially putting a… | Terminated |
| Apr 25, 2006 | Z-1115-06Improper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections. | Terminated |
| Jun 30, 2005 | Z-0889-05Image reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder. | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 27, 2000
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files