Arca
Skip to content
Tools
Tools

Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models)

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K001250, from Asahi Medical Co., Ltd. (Rockville, MD). Product code KDI, class 2.

As of Nov 21, 2026, K001250 is an FDA 510(k) held by Asahi Medical Co., Ltd., substantially equivalent on Aug 16, 2000 after 119 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 119 days
  • Recalls: 2
  • Cited as a predicate by: 8 devices

Cited as a predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K001250, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 31, 2005Z-0790-05Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.Terminated
Oct 1, 2001Z-0739-04Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Apr 19, 2000
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files