Arca
Skip to content
Tools
Tools

Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC Universal

System, Balloon, Intra-Aortic And Control. 510(k) K000729, from Arrow Intl., Inc. (Everett, MA). Product code DSP, class 2.

As of Nov 24, 2026, K000729 is an FDA 510(k) held by Arrow Intl., Inc., substantially equivalent on May 19, 2000 after 74 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 74 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Arrow Ultra 8 Intra-Aortic Balloon Catheters (IAB) 8 Fr 30cc and 40cc Universal is clinically indicated for the following conditions: a. Acute Coronary Syndrome b. Cardiac and Non-Cardiac Surgery c. Complications of Heart Failure

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K000729, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 29, 2024Z-1879-2024, Z-1880-2024Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon…Open
Feb 10, 2016Z-1060-2016, Z-1059-2016, Z-1062-2016, Z-1058-2016, Z-1056-2016The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.Terminated
Nov 27, 2012Z-1065-2013, Z-1064-2013Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.Terminated
Oct 11, 2010Z-0791-2011, Z-0790-2011, Z-0969-2011, Z-0793-2011, Z-0792-2011SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.Terminated
Feb 2, 2009Z-1216-2009, Z-1219-2009, Z-1218-2009, Z-1214-2009, Z-1221-2009, Z-1210-2009, Z-1215-2009, Z-1220-2009, Z-1217-2009, Z-1211-2009, Z-1212-2009A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.Terminated

Other DSP clearances

The newest 510(k)s for the same product code; see every DSP device

Record

Type
Special
Received
Mar 6, 2000
Review panel
Cardiovascular
Regulation
21 CFR 870.3535

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc