VENTANA Kappa and Lambda Dual ISH mRNA Probe Cocktail (08507023001)
Kappa And Lambda Immunoglobulin Light Chain In Situ Hybridization Mrna Probe Detection Kit. De Novo DEN240025, from Ventana Medical Systems, Inc. (Tucson, AZ). Product code SDP, class 2.
As of Nov 29, 2026, DEN240025 is an FDA De Novo held by Ventana Medical Systems, Inc., granted on Dec 5, 2024 after 190 days, with no recalls tied to it in FDA's recall database.
- Review time: 190 days
- Recalls: 0
The VENTANA Kappa and Lambda Dual ISH mRNA Probe Cocktail Assay is a qualitative assay that is used to detect the expression of kappa and lambda immunoglobulin light chains in formalin- fixed paraffin embedded (FFPE) human hematolymphoid specimens by in situ hybridization (ISH). The assay is intended as an aid in the diagnosis of mature B-cell lymphomas and plasma cell neoplasms.
- Design
- Multicenter method comparison study of archived bone marrow/lymphoid tissue samples vs. historical flow cytometry
- Enrolled
- 742
- Where
- 3 sites
- Primary endpoint
- 95% CI lower bounds of KRPA, LRPA, and NRPA agreement with historical flow cytometry restriction status (co-primary); lower bound must be at least 85%
- Result
- Met: KRPA 90.9% (CI lower 87.9%), LRPA 94.8% (CI lower 92.0%), NRPA 95.2% (CI lower 93.3%), all lower bounds above 85%
- Japan approved Ventana DP200 Dx System, the same brandPMDA · Jun 5, 2025
- Type
- Direct
- Received
- May 29, 2024
- Review panel
- Hematology
- Regulation
- 21 CFR 864.1861
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files