Versius Surgical System
Modular Electromechanical Surgical System. De Novo DEN230078, from Cmr Surgical Limited (Cambridge, GB). Product code SCV, class 2.
As of Nov 22, 2026, DEN230078 is an FDA De Novo held by Cmr Surgical Limited, granted on Oct 11, 2024 after 325 days, with no recalls tied to it in FDA's recall database.
- Review time: 325 days
- Recalls: 0
- Cited as a predicate by: 3 devices
The Versius Surgical System is a robotically assisted surgical device that is intended to assist in the precise and accurate control of Versius Surgical endoscopic instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, electrosurgery, and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and…
- Design
- Two prospective, single-center, single-arm studies (Poland, India) of 30 subjects each, robotic cholecystectomy
- Enrolled
- 60
- Where
- 2 sites, Poland, India
- Primary endpoint
- Rate of unplanned conversion to other minimal access or open surgery; primary safety: rate of intraoperative and postoperative SAEs to 30 days
- Result
- Poland: 29/30 (96.6%) completed without unplanned conversion, 2 postoperative SAEs in converted patient; India: 30/30 (100%) completed as planned, 0 SAEs
- Anovo Surgical System (6Ne)K260804 · Momentis Surgical , Ltd. · Jul 22, 2026
- Versius Surgical System (Versius Plus)K252111 · Cmr Surgical Limited · Dec 16, 2025
- Hugo™ RAS SystemK250725 · Covidien, LLC · Dec 3, 2025
- Versius Surgical System (Versius Plus)K261252 · Cmr Surgical Limited · Oct 21, 2026
- Anovo Surgical System (6Ne)K260804 · Momentis Surgical , Ltd. · Jul 22, 2026
- Versius Surgical System (Versius Plus)K252111 · Cmr Surgical Limited · Dec 16, 2025
- Hugo™ RAS SystemK250725 · Covidien, LLC · Dec 3, 2025
- Type
- Direct
- Received
- Nov 21, 2023
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4964
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files