Notal Vision Home Optical Coherence Tomography (OCT) System
Home Monitoring Ophthalmic Optical Coherence Tomography (Oct) Imaging Device. De Novo DEN230043, from Notal Vision, Inc. (Manassas, VA). Product code SAX, class 2.
As of Nov 21, 2026, DEN230043 is an FDA De Novo held by Notal Vision, Inc., granted on May 15, 2024 after 343 days, with no recalls tied to it in FDA's recall database.
- Review time: 343 days
- Recalls: 0
The Notal Vision Home Optical Coherence Tomography (OCT) System is an Artificial Intelligence (AI)-based Home Use device indicated for visualization of intraretinal and subretinal hypo-reflective spaces in a 10 by 10-degrees area centered on the point of fixation of eyes diagnosed with neovascular age-related macular degeneration (NV-AMD).
- Design
- Prospective, longitudinal, multicenter at-home study vs. RC-graded in-office CIRRUS HD-OCT reference
- Enrolled
- 198
- Where
- 7 sites, US
- Primary endpoint
- PPA and NPA of central macular total retinal fluid (TRO) on NVHO vs. RC-graded CIRRUS OCT; success rates of NVHO setup and self-imaging
- Result
- PPA 0.864 (95% CI 0.802-0.926; p=0.043); NPA 0.849 (95% CI 0.792-0.907; p=0.094); setup success 86.7%; self-imaging success 96.1%
- On FDA's list of AI-enabled devicesFDA AI-enabled medical devices
- Type
- Direct
- Received
- Jun 7, 2023
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files