Arca
Skip to content
Tools
Tools

Dream Sock

Infant Pulse Rate And Oxygen Saturation Monitor For Over-The-Counter Use. De Novo DEN220091, from Owlet Babycare, Inc. (Lehi, UT). Product code QYU, class 2.

As of Nov 18, 2026, DEN220091 is an FDA De Novo held by Owlet Babycare, Inc., granted on Nov 8, 2023 after 329 days, with no recalls tied to it in FDA's recall database.

  • Granted: Nov 8, 2023
  • Review time: 329 days
  • Recalls: 0
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Dream Sock analyzes photoplethysmography data to identify instances when the infant’s pulse rate (PR) and/or oxygen saturation (SpO2) moves outside a preset range, and provides a notification to the caregiver, prompting them to assess the infant.

Pivotal Study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Clinical at-home study in 35 subjects independent of the algorithm training set, accuracy vs. reference pulse oximeter
Enrolled
35
Primary endpoint
Accuracy (RMS) of SpO2 and pulse rate in at-home use
Result
SpO2 average A_rms 2.16%; pulse rate average A_rms 3.53 bpm

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • Masimo StorkK234021 · Masimo Corporation · May 3, 2024

Other QYU Clearances

The newest 510(k)s for the same product code; see every QYU device
  • Masimo StorkK234021 · Masimo Corporation · May 3, 2024

Record

Type
Direct
Received
Dec 14, 2022
Review panel
Cardiovascular
Regulation
21 CFR 870.2705

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.