Sunrise Sleep Disorder Diagnostic Aid
Device For Sleep Apnea Testing Based On Mandibular Movement. De Novo DEN210015, from Sunrise (Namur, BE). Product code QRS, class 2.
As of Nov 18, 2026, DEN210015 is an FDA De Novo held by Sunrise, granted on Jan 7, 2022 after 280 days, with no recalls tied to it in FDA's recall database.
- Granted: Jan 7, 2022
- Review time: 280 days
- Recalls: 0
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Sunrise SDDA device is a non-invasive home care aid in the evaluation of obstructive sleep apnea (OSA) in patients 18 years and older with suspicions of sleep breathing disorders. FDA concludes that this device should be classified into Class II.
Pivotal Study
The main clinical evidence FDA relied on, as its decision summary describes it- Design
- Prospective, comparative, open, independent study in France vs. ambulatory at-home PSG (NOX A1)
- Where
- France
- Primary endpoint
- Agreement of sleep parameters with PSG: OSA severity sensitivity/specificity at ORDI cut-offs and RMSE for sleep structure (TST, SOL, WASO, SE, ArI)
- Result
- Sensitivity and specificity at three ORDI cut-offs and RMSEs reported, but the numeric values are redacted in the source text
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- SunriseK222262 · Sunrise SA · Dec 22, 2022
Other QRS Clearances
The newest 510(k)s for the same product code; see every QRS deviceRecord
- Type
- Direct
- Received
- Apr 2, 2021
- Review panel
- Ear, Nose, Throat
- Regulation
- 21 CFR 868.2376
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files