Gili Pro BioSensor (also known as “Gili BioSensor System”)
Hardware And Software For Optical Camera-Based Measurement Of Heart Rate And Respiratory Rate. De Novo DEN200038, from Continuse Biometrics , Ltd. (Tel-Aviv, IL). Product code QOK, class 2.
As of Nov 23, 2026, DEN200038 is an FDA De Novo held by Continuse Biometrics , Ltd., granted on Apr 1, 2021 after 293 days, with no recalls tied to it in FDA's recall database.
- Review time: 293 days
- Recalls: 0
- Cited as a predicate by: 1 device
The Gili Pro BioSensor (Gili BioSensor System) includes an optical module that is intended to capture motion-vibration signals from an illuminated surface for assessment of physiological information. Such information, captured during spot-measurement, includes: • Heart rate • Respiratory rate The device is indicated for use by or under the supervision of healthcare professionals for adult…
- Design
- Two-phase clinical study: 10-subject pilot phase and 120-subject main pivotal study vs. ECG and capnography reference
- Enrolled
- 120
- Primary endpoint
- Co-primary: HR accuracy vs. ECG in non-arrhythmia subjects (±3 bpm); RR accuracy vs. over-scored capnography (±2 BrPM)
- Result
- Met: HR intercept 0.51 (95% CI 0.18, 0.84) within ±3 bpm; RR intercept 0.17 (95% CI -0.08, 0.42) within ±2 BrPM
- Echo-2K254240 · Neko Health AB · Sep 22, 2026
- Echo-2K254240 · Neko Health AB · Sep 22, 2026
- On FDA's list of AI-enabled devicesFDA AI-enabled medical devices
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Jun 12, 2020
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2786
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files