Arca
Skip to content
Tools
Tools

ClearUP Sinus Relief

Transcutaneous Electrical Nerve Stimulator For The Relief Of Congestion. De Novo DEN200006, from Tivic Health Systems, Inc. (Menlo Park, CA). Product code QNU, class 2.

As of Nov 22, 2026, DEN200006 is an FDA De Novo held by Tivic Health Systems, Inc., granted on Mar 5, 2021 after 392 days, with no recalls tied to it in FDA's recall database.

  • Review time: 392 days
  • Recalls: 0

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

ClearUP Sinus Relief device is a transcutaneous electrical nerve stimulator that electrically stimulates the skin overlying the paranasal sinuses and is intended to be used for the temporary relief of moderate to severe congestion. ClearUP Sinus Relief is a treatment to be used at home by individuals 18 and older. FDA concludes that this device should be classified into Class II.

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, double-blind, randomized, sham-controlled trial (Congestion Study 1)
Enrolled
71
Primary endpoint
Decrease in pain score active vs. sham; also change in Modified NOSE congestion score active vs. sham
Result
Congestion difference vs. sham not significant in all 71 subjects; significant only in post hoc moderate-to-severe subgroup; pain result not reported

Record

Type
Direct
Received
Feb 7, 2020
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.6000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files