SafeBreak Vascular
Intravenous Catheter Force-Activated Separation Device.. De Novo DEN190043, from Site Saver, Inc. D/B/A Lineus Medical (Fayetteville, AR). Product code QOI, class 2.
As of Nov 20, 2026, DEN190043 is an FDA De Novo held by Site Saver, Inc. D/B/A Lineus Medical, granted on May 27, 2021 after 617 days, with no recalls tied to it in FDA's recall database.
- Review time: 617 days
- Recalls: 0
- Cited as a predicate by: 3 devices
SafeBreak® Vascular is intended to separate when excessive tension is exerted across a peripheral IV administration set. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient’s IV catheter. SafeBreak® Vascular is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement.
- Design
- Prospective, randomized, controlled non-inferiority trial vs. standard of care (Control) at one hospital
- Where
- 1 site
- Primary endpoint
- Total delay of therapy per 24h per subject, SafeBreak vs. Control, non-inferiority; primary safety: delays and adverse events
- Result
- Met: UCL of CI for treatment difference within NI margin, non-inferior in ITT and PP populations (values redacted); 139 PP subjects (70 SafeBreak, 69 Control)
- SafeBreak® VascularK231957 · Lineus Medical · Sep 29, 2023
- SafeBreak VascularK222791 · Lineus Medical, LLC · Oct 16, 2022
- Orchid Safety Release Valve(TM)K212064 · Linear Health Sciences, LLC · May 3, 2022
- CleanBreak™K260478 · Icu Medical, Inc. · Nov 6, 2026
- Orchid Safety Release Valve™K241415 · Linear Health Sciences, LLC · Sep 6, 2024
- Orchid Safety Release ValveTMK232094 · Linear Health Sciences, LLC · Oct 3, 2023
- SafeBreak® VascularK231957 · Lineus Medical · Sep 29, 2023
- SafeBreak® VascularK223486 · Lineus Medical · May 19, 2023
- Orchid Safety Release ValveK230266 · Linear Health Sciences, LLC · May 4, 2023
- SafeBreak VascularK222791 · Lineus Medical, LLC · Oct 16, 2022
- Orchid Safety Release Valve(TM)K212064 · Linear Health Sciences, LLC · May 3, 2022
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Sep 18, 2019
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5220
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files