Caption Guidance
Image Acquisition And/Or Optimization Guided By Artificial Intelligence. De Novo DEN190040, from Bay Labs, Inc. (San Francisco, CA). Product code QJU, class 2.
As of Nov 20, 2026, DEN190040 is an FDA De Novo held by Bay Labs, Inc., granted on Feb 7, 2020 after 164 days, with no recalls tied to it in FDA's recall database.
- Review time: 164 days
- Recalls: 0
- Cited as a predicate by: 6 devices
The Caption Guidance software is intended to assist medical professionals in the acquisition of cardiac ultrasound images. Caption Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems.
- Design
- Prospective pivotal study of RN-acquired exams with Caption Guidance vs. sonographer control exams, read by cardiologist panel
- Primary endpoint
- Four endpoints: proportion of RN study exams with image quality sufficient for qualitative visual assessment of LV size, LV function, RV size, and non-trivial pericardial effusion
- Result
- Met: 98.8% (LV size), 98.8% (LV function), 92.5% (RV size), 98.8% (pericardial effusion) of scans sufficient quality; met predetermined success criteria
- ThinkSono GuidanceK260338 · ThinkSono GmbH · Jul 15, 2026
- Cardiac GuidanceK243065 · Caption Health, Inc. · Jan 15, 2025
- HeartFocus (V.1.1.1)K242807 · Deski · Apr 4, 2025
- AI Platform 2.0 (AIP002)K240953 · Exo Imaging · Aug 5, 2024
- UltraSight AI GuidanceK223347 · Ultrasight, Inc. · Jul 24, 2023
- Caption GuidanceK200755 · Caption Health · Apr 16, 2020
- Lung GuidanceK261042 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · Oct 2, 2026
- ThinkSono GuidanceK260338 · ThinkSono GmbH · Jul 15, 2026
- HeartFocusK260780 · Deski · Jun 3, 2026
- UltraSight GuidanceK251416 · Ultrasight , Ltd. · Dec 17, 2025
- HeartFocus (V.1.1.1)K242807 · Deski · Apr 4, 2025
- Cardiac GuidanceK243065 · Caption Health, Inc. · Jan 15, 2025
- UltraSight AI GuidanceK223347 · Ultrasight, Inc. · Jul 24, 2023
- On FDA's list of AI-enabled devicesFDA AI-enabled medical devices
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Aug 27, 2019
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2100
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files