Acumen Assisted Fluid Management (AFM) Software Feature
Adjunctive Open Loop Fluid Therapy Recommender. De Novo DEN190029, from Edwards Lifesciences (Irvine, CA). Product code QMS, class 2.
As of Nov 18, 2026, DEN190029 is an FDA De Novo held by Edwards Lifesciences, granted on Nov 13, 2020 after 529 days, with no recalls tied to it in FDA's recall database.
- Review time: 529 days
- Recalls: 0
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Edwards Lifesciences Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring.
Pivotal study
The main clinical evidence FDA relied on, as its decision summary describes it- Design
- Prospective, multi-center, single-arm study vs. historical performance criterion of 30% (AFM IDE study)
- Enrolled
- 330
- Where
- US
- Primary endpoint
- Percentage of AFM Recommendation-prompted boluses that increased stroke volume per the selected fluid strategy; primary safety endpoint: device-related SAEs
- Result
- Met: 66.1% [62.1, 69.7] of AFM Recommendation boluses increased SV per fluid strategy vs. 30% historical criterion; no ADEs/SADEs adjudicated definitely related to device
Cited as a predicate
1 later 510(k) summary names this device as a predicate- HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable,…K213682 · Edwards Lifesciences, LLC · Jun 22, 2022
Other QMS clearances
The newest 510(k)s for the same product code; see every QMS device- Acumen Assisted Fluid Management (AFM) Software FeatureK233984 · Edwards Lifesciences, LLC · Aug 2, 2024
Record
- Type
- Direct
- Received
- Jun 3, 2019
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files