Arca
Skip to content
Tools
Tools

ECG App

Electrocardiograph Software For Over-The-Counter Use. De Novo DEN180044, from Apple, Inc. (Alexandria, VA). Product code QDA, class 2.

As of Nov 21, 2026, DEN180044 is an FDA De Novo held by Apple, Inc., granted on Sep 11, 2018 after 28 days, with no recalls tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 0
  • Cited as a predicate by: 6 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ECG app is a software-only mobile medical application intended for use with the Apple Watch to create, record, store, transfer, and display a single channel electrocardiogram (ECG) similar to a Lead I ECG. The ECG app determines the presence of atrial fibrillation (AFib) or sinus rhythm on a classifiable waveform. The ECG app is not recommended for users with other known arrhythmias.

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, parallel-cohort, non-randomized, multi-center reader study using an enriched population
Enrolled
602
Where
5 sites
Primary endpoint
Sensitivity and specificity of AF detection by ECG App algorithm vs. physician-adjudicated 12-lead ECG; performance goals 90% and 92%
Result
Met: AF sensitivity 98.3% (97.5% LCB 95.8%) and specificity 99.6% (97.5% LCB 97.7%) in classifiable set

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Other QDA clearances

The newest 510(k)s for the same product code; their median review was 176 days; see every QDA device

Record

Type
Direct
Received
Aug 14, 2018
Review panel
Cardiovascular
Regulation
21 CFR 870.2345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files