Irregular Rhythm Notification Feature
Photoplethysmograph Analysis Software For Over-The-Counter Use. De Novo DEN180042, from Apple, Inc. (Alexandria, VA). Product code QDB, class 2.
As of Nov 19, 2026, DEN180042 is an FDA De Novo held by Apple, Inc., granted on Sep 11, 2018 after 33 days, with no recalls tied to it in FDA's recall database.
- Review time: 33 days
- Recalls: 0
- Cited as a predicate by: 4 devices
The Irregular Rhythm Notification Feature is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for
- Design
- Prospective, single-arm sub-study of participants with one prior notification; PPV vs. 7-day ambulatory ECG patch
- Enrolled
- 269
- Primary endpoint
- PPV of spot irregular tachograms to detect AF vs. ambulatory ECG patch monitoring reference, vs. pre-specified performance goal
- Result
- Not met: PPV 66.6% (lower 97.5% confidence bound 63.0%) failed pre-specified performance goal
- Samsung ECG Monitor Application with Irregular Heart Rhythm NotificationK230292 · Samsung Electronics Co., Ltd. · May 2, 2023
- Rune Labs Tremor Transducer SystemK213519 · Rune Labs, Inc. · Jun 10, 2022
- IRNF AppK212516 · Apple, Inc. · Oct 22, 2021
- Fitbit Irregular Rhythm NotificationsK212372 · Fitbit, Inc. · Apr 8, 2022
- Irregular Rhythm Notification Feature (IRNF)K231173 · Apple, Inc. · Jul 21, 2023
- Atrial Fibrillation History FeatureK213971 · Apple, Inc. · Jun 3, 2022
- Fitbit Irregular Rhythm NotificationsK212372 · Fitbit, Inc. · Apr 8, 2022
- IRNF AppK212516 · Apple, Inc. · Oct 22, 2021
- Type
- Direct
- Received
- Aug 9, 2018
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2790
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files