Quantra Hemostasis Analyzer, QPlus Cartridge

Coagulation System For The Measurement Of Whole Blood Viscoelastic Properties. De Novo DEN180017, from Hemosonics, LLC (Charlottesville, VA). Product code QFR, class 2.

As of Nov 21, 2026, DEN180017 is an FDA De Novo held by Hemosonics, LLC, granted on Mar 12, 2019 after 344 days, with no recalls tied to it in FDA's recall database.

  • Review time: 344 days
  • Recalls: 0
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Quantra QPlus System is composed of the Quantra Hemostasis Analyzer, QPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra QPlus System is intended for in vitro diagnostic use. The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation.

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, multi-center, observational clinical performance study in perioperative cardiac, vascular, and orthopedic surgery patients vs. standard laboratory tests and ROTEM
Enrolled
302
Where
4 sites, US
Primary endpoint
Overall agreement of each QPlus parameter's classification (Low/Normal/High) with clinical composite index (CCI) of standard laboratory tests; also concordance vs. ROTEM delta parameters
Result
Overall agreement: CT vs CCI1 0.72 (lower bound 0.69); CS vs CCI2 0.77; FCS vs CCI3 0.84; PCS vs CCI4 0.75; CTR vs CCI5 0.98 (lower bound CIs reported; 277 eligible subjects)

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Other QFR clearances

The newest 510(k)s for the same product code; their median review was 82 days; see every QFR device

Record

Type
Direct
Received
Apr 2, 2018
Review panel
Hematology
Regulation
21 CFR 864.5430

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files