Radiological Computer-Assisted Triage And Notification Software. De Novo DEN170073, from Viz.Al, Inc. (San Francisco, CA). Product code QAS, class 2.

As of Nov 19, 2026, DEN170073 is an FDA De Novo held by Viz.Al, Inc., granted on Feb 13, 2018 after 137 days, with no recalls tied to it in FDA's recall database.

  • Review time: 137 days
  • Recalls: 0
  • Cited as a predicate by: 11 devices

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Retrospective study of 300 CTA cases from two US clinical sites vs. neuro-radiologist ground truth
Enrolled
300
Where
2 sites, US
Primary endpoint
Sensitivity and specificity of ContaCT LVO notification vs. ground truth and pre-defined performance goals
Result
Sensitivity 87.8% (95% CI 81.2-92.5%) and specificity 89.6% (95% CI 83.7-93.9%); primary endpoints exceeded performance goals

Cited as a predicate

11 later 510(k) summaries name this device as a predicate
  • Annalise EnterpriseK253818 · Harrison-AI Medical Pty, Ltd. · Mar 3, 2026
  • Methinks CTA StrokeK251590 · Methinks Software, S.L · Aug 20, 2025
  • Viz LVO ContaCTK223042 · Viz. Ai, Inc. · Oct 21, 2022
  • StrokeSENS LVOK212261 · Circle Neurovascular Imaging, Inc. · Oct 14, 2021
  • HaloK200873 · Nico-Lab B.V. · Nov 20, 2020
  • Viz ICHK193658 · Viz. Ai, Inc. · Mar 18, 2020
  • BriefCaseK192383 · Aidoc Medical , Ltd. · Dec 20, 2019
  • cmTriageK183285 · Curemetrix, Inc. · Mar 8, 2019
  • DeepCTK182875 · Deep01 Limited · Jul 10, 2019
  • AccipiolxK182177 · Maxq-Al , Ltd. · Oct 26, 2018
  • BriefCaseK180647 · Aidoc Medical , Ltd. · Aug 1, 2018

Other QAS clearances

The newest 510(k)s for the same product code; their median review was 104 days; see every QAS device

Elsewhere

Record

Type
Direct
Received
Sep 29, 2017
Review panel
Radiology
Regulation
21 CFR 892.2080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files