QuantX
Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer. De Novo DEN170022, from Quantitative Insights, Inc. (Chicago, IL). Product code POK, class 2.
As of Nov 20, 2026, DEN170022 is an FDA De Novo held by Quantitative Insights, Inc., granted on Jul 19, 2017 after 103 days, with no recalls tied to it in FDA's recall database.
- Review time: 103 days
- Recalls: 0
- Cited as a predicate by: 4 devices
- Design
- Retrospective, multiple reader multiple case (MRMC) sequential-read study of breast MR cases, second read with QuantX vs. first read without
- Enrolled
- 111
- Where
- 3 sites
- Primary endpoint
- Difference in AUC between second read (with QuantX) and first read (without QuantX) for breast cancer diagnosis, H1: dAUC > 0
- Result
- Proper-binormal dAUC 0.0520 (95% CI 0.0022 to 0.1018, p=0.0408); trapezoidal dAUC 0.0487 (95% CI -0.0011 to 0.0985, p=0.0550); study marginally met primary endpoint
- Avenda Health AI Prostate Cancer Planning SoftwareK221624 · Avenda Health, Inc. · Nov 22, 2022
- Optellum Virtual Nodule Clinic, Optellum software, Optellum platformK202300 · Optellum, Ltd. · Mar 5, 2021
- Rapid ASPECTSK200760 · Ischemaview, Inc. · Jun 26, 2020
- Koios DS for BreastK190442 · Koios Medical, Inc. · Jul 3, 2019
- BlineSlideK261679 · Deep Breathe, Inc. · Oct 23, 2026
- Koios DS (3.8)K253884 · Koios Medical, Inc. · Jul 2, 2026
- Sonio SuspectK261519 · Sonio · Jun 4, 2026
- Automated Aortic Stenosis Software (AutoAS)K254161 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · Mar 27, 2026
- RevealAI-LungK251769 · Precision Medical Ventures, Inc. Dba Revealdx · Jan 30, 2026
- Fetal EchoScan (v1.2)K252294 · Brightheart · Dec 8, 2025
- StrokeSENS ASPECTS Software ApplicationK250221 · Circle Cardiovascular Imaging, Inc. · Jul 1, 2025
- Fetal EchoScan (v1.1)K251071 · Brightheart · May 2, 2025
- On FDA's list of AI-enabled devicesFDA AI-enabled medical devices
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Apr 7, 2017
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2060
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files