Arca
Skip to content
Tools
Tools

Bonebridge

Active Implantable Bone Conduction Hearing System. De Novo DEN170009, from Med-El Elektromedizinische Geraete GmbH (Innsbruck, AT). Product code PFO, class 2.

As of Nov 23, 2026, DEN170009 is an FDA De Novo held by Med-El Elektromedizinische Geraete GmbH, granted on Jul 20, 2018 after 522 days, with no recalls tied to it in FDA's recall database.

  • Review time: 522 days
  • Recalls: 0
  • Cited as a predicate by: 4 devices

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, single-arm, open-label study (BB001) of adult and pediatric patients with mixed or conductive hearing loss, with 12-month follow-up
Enrolled
24
Primary endpoint
Improvement in word/sentence recognition from preoperative unaided to 12-month postoperative aided condition (success criterion 15%); primary safety endpoint: AEs and SAEs through 12 months
Result
Met: average 63.3% improvement in Freiburger word recognition at 12 months (p<0.001; 62.9% adults, 65.2% children) vs. 15% success criterion

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Other PFO clearances

The newest 510(k)s for the same product code; their median review was 114 days; see every PFO device

Record

Type
Direct
Received
Feb 13, 2017
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.3340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files