Acumen Hypotension Prediction Index (HPI) Feature Software
Adjunctive Predictive Cardiovascular Indicator. De Novo DEN160044, from Edwards Lifesciences, LLC (Irvine, CA). Product code QAQ, class 2.
As of Nov 17, 2026, DEN160044 is an FDA De Novo held by Edwards Lifesciences, LLC, granted on Mar 16, 2018 after 536 days, with no recalls tied to it in FDA's recall database.
- Granted: Mar 16, 2018
- Review time: 536 days
- Recalls: 0
- Cited as a predicate by: 6 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient’s likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics.
Pivotal Study
The main clinical evidence FDA relied on, as its decision summary describes it- Design
- Prospective, protocol-collected multicenter clinical validation database of acute care OR patients vs. annotated hypotensive events
- Enrolled
- 52
- Primary endpoint
- Sensitivity, specificity, and AUC for predicting hypotension (MAP <65 mmHg for at least 1 min) at HPI threshold 85
- Result
- Sensitivity 83.7% [81.5, 86.0], specificity 99.8% [99.4, 100.0], AUC 0.95 (N=52); no stated performance goal
Cited as a Predicate
6 later 510(k) summaries name this device as a predicate- Acumen Hypotension Prediction Index (HPI) AlgorithmK230057 · Edwards Lifesciences, LLC · Jun 8, 2023
- HemoSphere Advanced Monitor, HemoSphere ClearSight Module, Acumen Hypotension Prediction…K203687 · Edwards Lifesciences, LLC · May 28, 2021
- Acumen Hypotension Prediction Index, HemoSphere Advanced Monitoring PlatformK203224 · Edwards Lifesciences, LLC · Jul 30, 2021
- CLEWICU System (ClewICUServer and ClewICUnitor)K200717 · Clew Medical , Ltd. · Jan 9, 2021
- Acumen Hypotension Prediction Index – EV1000 Clinical Platform, Acumen Hypotension…K183646 · Edwards Lifeciences, LLC · May 21, 2019
- HemoSphere Advanced Monitor, HemoSphere Pressure CableK180881 · Edwards Lifesciences, LLC · Nov 16, 2018
Other QAQ Clearances
The newest 510(k)s for the same product code; their median review was 265 days; see every QAQ device- Hypertension Prediction Index (HePI) AlgorithmK242518 · Edwards Lifesciences, LLC · May 15, 2025
- Acumen Hypotension Prediction Index (HPI) AlgorithmK230057 · Edwards Lifesciences, LLC · Jun 8, 2023
- Acumen Hypotension Prediction Index, HemoSphere Advanced Monitoring PlatformK203224 · Edwards Lifesciences, LLC · Jul 30, 2021
Elsewhere
- On FDA's list of AI-enabled devicesFDA AI-enabled medical devices
Record
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Sep 26, 2016
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2210
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files