Arca
Skip to content
Tools
Tools

SkinPen Precision System

Powered Microneedle Device. De Novo DEN160029, from Bellus Medical, LLC (Dallas, TX). Product code QAI, class 2.

As of Nov 19, 2026, DEN160029 is an FDA De Novo held by Bellus Medical, LLC, granted on Mar 1, 2018 after 604 days, with no recalls tied to it in FDA's recall database.

  • Review time: 604 days
  • Recalls: 0
  • Cited as a predicate by: 10 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

SkinPen® Precision System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older. FDA concludes that this device should be classified into Class II.

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, single-center study; blinded photo grading of acne scars vs. baseline, SkinPen group n=20 analyzed
Enrolled
65
Where
1 site
Primary endpoint
Change in Acne Scar Assessment Scale score from baseline at 6 months (blinded dermatologist photo grading); CGAIS also listed as primary
Result
Mean score 2.80 at baseline vs. 2.35 at 6 months; 55% (11/20) of SkinPen subjects improved; no subject worsened; no pre-specified success threshold stated

Cited as a predicate

10 later 510(k) summaries name this device as a predicate

Other QAI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every QAI device

Record

Type
Direct
Received
Jul 5, 2016
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4430

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files